Light
Scattering Immunoassays
•Uwhen ag-ab
complexes are formed in a solution they form immunoprecipitates which have the ability to
scatter a light beam
that passes through the solution.
•Turbidimetric
application:
•U detectors
set at 180o angle measures reduction in light
•Nephelometric
application:
•U detectors
set at 90o angle measures increases in light
•in both
techniques standard curves are constructed to convert absorbance to concentration
•in
turbidimetry the absorbance is directly proportionate to concentration so std curve is linear and fewer standards needed
•in
nephelimetry the standard curve is non-linear and requires calibrators